9  Specications
Analog outputs
2 or 4 independent outputs
0 22 mA configurable as:
` 0 20 mA
` 4 20 mA
` 22 mA as alarm signal
Analog output accuracy ±0.2 % f.s.
Digital
communication
RS‑485 serial port for remote monitoring and control (Modbus)
USB‑C port to retrieve log files and firmware upgrading
Relays
Up to 5 relays, independently configurable for process variables, Hold and Cleaning
functions
Electromechanical relay outputs
5 A 250 Vac; 5 A 30 Vdc (resistive load)
Fuse protected: 5 A, 250 V slow blow fuse
Alarm relay for all
measurement alarms
Electromechanical relay SPDT contact output
5 A 250 Vac; 5 A 30 Vdc (resistive load)
Fuse protected: 5 A, 250 V slow blow fuse
Data logging
Interval log, up to 100 files, maximum 8600 records on each stored file.
When the maximum limit is reached, the most recent file will automatically erase
the oldest one.
Event log, maximum 100 records.
When the maximum limit is reached, the last record overwrites the oldest one.
Power supply
100 240 Vac ±10 %; 50/60 Hz; 15 VA;
fuse protected (2 A, 250 V slow blow fuse)
Power consumption 15 VA
Installation category II
Enclosure* Single case ½ DIN, IP65 ingress protection
Weight Approximately 1.6 kg (3.5 lb.)
Dimensions
Width 144.0 mm (5.7”)
Height 144.0 mm (5.7”)
Depth 151.3 mm (6.0”)
Environment –20 to 50 °C (–4 to 122 °F); maximum 100 % RH non‑condensing
* For a water tight seal, tighten the four front casing screws to 13.3 lbfin (1.5 Nm, max. 2.0 Nm), of torque.