Configuring the CX22B‒W for self-contained measurements
64 A3169-8.0 HBM: public CX22B / CX22B-W
Settings must be made in two areas to configure the
CX22B‒W for self-contained measurement:
S You must make the applicable settings in DAQ jobs
and options in catman
EASY so that catman
EASY
loads a DAQ project and launches the DAQ job after
startup.
S You must ensure that catman
EASY starts when the
CX22B‒W is switched on.
As usual, measurement can be started and stopped
immediately, or via a trigger. But you can also use the
START/STOP button of the CX22B‒W, see Section 4.9
on Page 21.
Procedure
Configure your DAQ project as described in Section
8.1. Set up the DAQ jobs so that the data are saved
with the Automatically on DAQ termination setting
(Data saving in the Settings group, Data storage
parameter).
Save the DAQ project.
Call up the catman
EASY options (file menu) and go
to Program start (System group).
For Start mode, select an Existing DAQ project and
choose the project created above in the DAQ project
line.
Activate the option Execute first DAQ job automati
cally.
Go to DAQ channels in the Channels and sensors
group.
Activate the option Automatically deactivate faulty
channels and continue measurement (behavior if
channel initialization fails).