Configuring the CX22B‒W for self-contained measurements
CX22B / CX22B-W A3169-8.0 HBM: public 65
► Exit the dialog by clicking on OK.
► Exit catman
EASY.
► In the start screen, click on Settings.
► On the General tab, activate the option Launch DAQ
software (catmanEASY) at CX22B‒W startup and
click on OK.
► Click on Commit changes and reboot to perma
nently save the new settings.
This project will then start automatically the next time the
CX22B‒W is switched on. If the power supply is inter
rupted during measurement, the project will be reloaded
and measurement continued when the power supply is
restored. The module restart, up to the continuation of
measurement, will require approx. 3 minutes, depending
on the number of connected devices and channels.
Important
The data already present in the catman
[
EASY tempo
rary measurement data memory are not lost if the power
supply is interrupted. Any additional files will be assigned
a counter at the end of the file name.