WARNING! DO NOT disconnect the driveline or power sources from the controller while
cleaning it or the pump will stop. If this happens, reconnect the driveline or power source to
the controller as soon as possible to restart the pump.
76
7.0 Caring for HeartWare
HVAD
System Equipment
7.1 General Care
The HeartWare
HVAD
System is made of durable materials that will need
occasional cleaning.
The following steps should be used to clean the equipment:
1. Use a clean, soft cloth when cleaning the system (controller, batteries,
battery charger).
2. Gently clean the equipment. Avoid hard rubbing.
7.2 Care of Your Controller
Once a week: Inspect the power connectors and connector pins on the controller
for dirt or grime. This inspection can be done when you are changing power sources.
Check the controller power connectors one at a time. DO NOT disconnect both
power sources at the same time – your pump will stop. DO NOT disconnect the
driveline to examine its connector. The only time the driveline connector should be
inspected is during a controller exchange. DO NOT attempt to clean the controller
connectors. If any dirt is found, report the condition to your clinician.
CAUTION: DO NOT attempt to repair or service any components of the HeartWare
HVAD
System. If service is required, contact your clinician.
CAUTION: ALWAYS keep all connectors free of liquid, dust and dirt, or the HeartWare
HVAD
System may not function as intended.
CAUTION: DO NOT expose the driveline to direct or indirect sunlight. Keep the driveline
completely covered when in the sun. DO NOT use tanning lights or black lights. The light from
these sources may damage the outer covering of the driveline.
WARNING! DO NOT use any components other than those supplied by HeartWare with the
HeartWare
HVAD
System, as this may affect HeartWare
HVAD
System operation.
WARNING! DO NOT disconnect the driveline or power sources from the controller while
cleaning it or the pump will stop. If this happens, reconnect the driveline to the controller as
soon as possible to restart the pump.
WARNING! DO NOT drop the controller or other equipment. Dropping the controller could
cause sudden stoppage of the pump. Dropped equipment should be reported and inspected.
WARNING! Damaged equipment should be reported to your clinician and inspected.