Service and Maintenance
6–4 Edition September 2000
ABC
6. Lift the defective fluorescent lamp out of
its mounting post.
Note: It is imperative that you only use clean
fluorescent lamps. If necessary, clean the lamp
before installation. Finger prints should likewise
be avoided as the grease can reduce the length of
operation.
7. Place the new fluorescent lamp into the mounting
post in such a way that it makes good contact.
8. Latch the black protective profile into the lamp
carrier.
9. Carefully reinsert the glass pane. Carefully slide
the pane against the stop pins to the rear and left.
Screw the 3 mounting posts into place. When
doing so make absolutely certain that the
mounting posts are inserted the right way round
to ensure that the glass pane does not break when
the screws are tightened.
10. Reconnect the device and switch it on.