Controlling E9Screen EPC9 Installation 89
acknowledged (re-synchronize command index)
syntax: "acknowledged"
no parameters
no response
ADRead
syntax: "ADRead AD-channel"
AD-channel: integer 0 to 7
returns a real (read AD-value in Volts)
ADRelease
syntax: "ADRelease"
no parameters
returns string "ADReleased"
ADReserve
syntax: "ADReserve"
no parameters
returns:
if it could reserve the AD-board: string "ADReserved"
otherwise: string "ADReleased"
DAWrite
syntax: "DAWrite AD-channel Volts"
DA-channel: integer, 0 to 3
Volts: real, +/-10.24 volt
no response
ExecMacro
syntax: "ExecMacro MacroIndex"
MacroIndex: integer, 0 to 6
no response
MeasureNoise
syntax: "MeasureNoise"
no parameters
returns a real
Query
syntax: "Query"
no parameters
returns the string "Running"
Reset
syntax: "Reset"
no parameters
no response
Terminate
syntax: "Terminate"
no parameters
returns the string "Terminated"
Key
syntax: "Key ASCII-code Key-Code"
Key-Codes:
2 -> Normal character input