6.8.4. Setting the time at which the alarm of disconnection or apnea
should be activated
The next available option within the Configuration
Mode (S) is setting the time at which the alarm of
disconnection or apnea should be activated (At).
Pressing the OK key, the letter t and below the
programmed time in seconds are displayed.
The default time is 30 seconds. The minimum is 10
seconds and the maximum 60 seconds.
In the
example 30 seconds are scheduled. This means the
alarm is activated when inspirations are not detected
within 30 seconds.
To modify value use Plus or Minus buttons.
Once selected, confirm by pressing OK.
I NOTE .
With Smart Assist activated, the alarm of disconnection or apnea will be
activated reached one of these two time limits: 1) the programmed time, or 2)
the Smart Assist system calculated time.