6.8.6. Configuration of automated Power-Off
The next available option inside the Configuration
Mode (S) is Auto power-Off (AO). Being this function
activated, when an alarm for apnea or disconnection
is not activated in 5 minutes, OXYMO interprets it is
not dealing with an apnea but a disconnection
(patient has taken off the nasal cannula forgetting to
switch off the unit). Provided this situation and to
avoid to waste battery, it turns off automatically.
Pressing OK button the option is activated or
deactivated.
Activated status is represented by an y and
deactivated by ann. In example it is activated.
To modify it, use Plus or Minus buttons.
Once configured, confirm pressing OK button.