6. OPERATION INSTRUCTIONS
Before connecting the device, check for damages that could avoid its correct
operation; if that is the case, you must contact the supplier or the technical
service.
6.1. CONNECTION TO AN O
2
CYLINDER
1
Check the oxygen cylinder (1) is closed.
Connect to cylinder an oxygen pressure
regulator (2) having a standard DISS
outlet pressure, of between 2.8 and 7
bar (in the figure, Hersill P-300). Open
the cylinder valve slowly. Check for
leaks and that the cylinder has enough
oxygen (the arrow points to
the green
area).
Connect the regulator OXYMO P-15
MICRO (3) to the cylinder regulator.
2
Connect the spiral hose to the regulator
OXYMO P-15 MICRO.
3
Connect the other terminal of the spiral
hose to the inlet connection of the unit.
Insert batteries (back door of the device)
and/or connect de AC/DC adapter
4
Turn-on the device (On button).
During the boot sequence the display
will blink three times and the buzzer will
beep.
I CAUTION .
Does not turn on OXYMO with nasal
cannula connected to the patient.