6.8.5. Configuration of Safety Flow for disconnection or apnea
Next available option inside the Configuration Mode
(S) is the activation of a security flow for apnea (AF).
Being this option activated, during an alarm for
apnea OXYMO supplies pulses of oxygen, reaching
a flow of 2 L/min, until the alarm is deactivated or the
device switched off. If the alarm is silenced (by the
keyboard), but inspirations still not detected, the
security flow is not interrupted.
Pressing OK button the option is activated or
deactivated.
Activated status is represented by an y and
deactivated by an n. In example it is activated.
To modify it, use Plus or Minus buttons.
Once configured, confirm pressing OK button.