7. ALARMS
OXYMO includes systems for detection of the main
events that affect the regular operation of device or
the regular behaviour of patient. To avoid risks, as
soon as one of these events is detected, an alarm
will be activated.
OXYMO will detect three different dysfunctions:
1. Low battery level.
2. Low O
2
pressure supply.
3. Disconnection or apnea.
Each alarm produces an audible warning and another visual indication on
screen (alarm symbols). When an alarm is activated, the acoustic warning
can be silenced pressing Silence alarm. After silencing the alarm, the
audible warning is definitively deactivated, but the visual warning on screen
will not stop blink until the problem gets solved.