13. TROUBLESHOOTING
Problems Possible causes Solutions
1.
There are oxygen
leaks
1. The spiral hose
is not well
connected
Make sure the spiral hose is
correctly connected to the outlet
of the pressure regulator and to
the inlet connector of OXYMO
(See Chapter 6)
2. The spiral hose
is broken or
damaged.
Replace the spiral hose for a new
one
3. The pressure
regulator is not
connected to the
cylinder
Close the cylinder and make sure
the regulator is fully inserted into
the cylinder
4. Pressure
regulator is faulty
Replace the pressure regulator for
a new one
2.
The unit won’t turn
on ( See Chapter
6.4)
1. It has no power
supply
Make sure that the batteries are
not exhausted and that they are
correctly positioned. If faulty,
replace them
Connect the adapter to the
network supply
2. The unit is
broken down
Disconnect the nasal cannula and
the spiral hose from OXYMO and
connect the adapter for a
continuous flow instead.
This way, a continuous oxygen
flow of 3 L/min will be achieved
Call the technical service or
oxygen provider