51
3.
Batteries will not
be charged (See
Chapter 8.1)
not correctly
Check connections
Replace the adapter for a new
one
batteries or they
are not
Insert rechargeable NiMH
batteries in the compartment for
batteries in OXYMO
4. Rechargeable
batteries are faulty
Replace the rechargeable
batteries for new ones
4.
Alarm for
disconnection/
apnea is activated
(See Chapter 7.3)
1. The unit does not
detect the patient
inspiration
Check the patient is not in apnea
Check the patient breaths through
the nostrils and not through the
Check the nasal cannula is
correctly located in the device, its
tubing not folded or clogged and it
is correctly placed in the nostrils
of the patient (See Chapter
6.4)
Do not use a nasal cannula more
than 2 meters long
Replace OXYMO for the 3 L/min
continuous flow accessory