52
5.
An alarm for low O
2
pressure supply is
activated (See
Chapter 7.2)
1. The oxygen
cylinder is empty
Replace the cylinder for a new
one
2. The oxygen
cylinder is closed
Open very slowly the valve of the
oxygen cylinder
3. The spiral hose
is clogged or folded
Remove the impediment or
crease or replace with a new unit
4. The oxygen
outlet connector in
OXYMO is clogged
Try to eliminate the dirt blocking
the oxygen flow. If not possible,
call the technical service or your
oxygen provider
6.
The device is
blocked( it does not
work, nor respond
to the keypad)
1. Power failure
while the device is
in operation
Disconnect the power supply
(battery, mains or USB) for more
than 30 seconds. This way the
unit is completely reset and it can
be boot up (date and hour shall
be reintroduced from the
Configuration Mode)