36
I NOTE .
Taking the equivalent flow to 0 L/min, OXYMO only performs monitor tasks
(Respiratory Monitor Mode). It does not release oxygen nor require be
connected to the oxygen supply.
6.9.4. Setting the trigger
During regular operation, the screen shows the
equivalent flow in L/min, the trigger symbol each time
that the bolus is on, the respiratory frequency and
the battery level.
To view or modify the trigger value, press Trigger
button. At that moment, it appears the current trigger
value.
In example, after pressing Trigger
showed on screen the current value of trigger, which
is from -1,5 mmH
2
O.
The trigger value can be modified between -1,5
mmH
2
O and -9,5 mmH
2
O. To modify its value, press
the Plus or Minus buttons. Each press
increase or decrease 0.5 mmH
2
O.
Once reached the desired trigger, it will blink for 5
seconds along with the OK symbol waiting for
confirmation pressing OK. If within those 5 seconds,
confirmation does not take place, the unit goes back
to previous status.