I NOTE .
Though not used, batteries get discharged slowly. If discharged, their
capacity for recharging is reduced. Recharge every 3 months, though not
used. OXYMO charge batteries in a progressive way and by cycles (quick,
slow and maintenance charging). This way NiMH batteries life and their
capacity are extended.
I CAUTION .
If NiMH batteries are not available, alkaline batteries can be used without any
damage to the device. The unit will identify them and will not get them
recharged. Charge level indicator is not to trust with alkaline batteries. In no
time it can go from indicating full charge to get an alarm on low battery.
8.2. CHANGE BATTERIES
To position batteries into the unit unscrew for accessing the opening in the
batteries cover. It is recommended to settle the screw again to get it difficult
for accidental or unauthorized withdrawal of the batteries.
I CAUTION .
Before replacing the batteries you must turn off the device
I CAUTION .
The keeping of dating and timing is a battery-related issue; if batteries are
withdrawn for more than 30 seconds, without being the unit connected to
power supply, date and time records will be lost and have to be settled again
(from the unit or from a PC). It is recommended to change batteries while
OXYMO is connected to power supply.