13/42
IMBALANCE
Error Imbalance on motor axle.
Error
consequence
The centrifuge slows down until the “open the lid” release
occurs.
Cause of error /
measurements
Weight difference in rotor components.
Supporting lugs not lubricated.
False IMBALANCE MODE is set (see chapter "Imbalance
Mode").
Imbalance switch not connected.
Imbalance switch is defective.
Measure on supply board / plug X3 (pin 4), see chapter
"Mains supply with supply board (A1)".
Loose contact in cable or plug.
Flat ribbon cable to control panel is defective.
Measure on control panel / plug X1 (pin 12)
control panel or supply board is defective.
Error code reset Open the lid or perform a MAINS RESET.
CONTROL - ERROR 04
Error Lid lock is open during centrifugation.
Error
consequence
Slowing down until the ”open the lid” release occurs.
Cause of error
Lid lock is defective and can be opened during
centrifugation.
Loose contact in cable or in plug.
Flat ribbon cable to control panel is defective.
Control panel or supply board is defective.
Measurement Measure on supply board / plug X5 (pin 6) and on control
panel / plug X1 (pin 5).
Error code reset Perform a MAINS RESET.

Table of Contents

Other manuals for Hettich EBA 21

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich EBA 21 and is the answer not in the manual?

Hettich EBA 21 Specifications

General IconGeneral
BrandHettich
ModelEBA 21
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals