13/42
IMBALANCE
Error Imbalance on motor axle.
Error
consequence
The centrifuge slows down until the “open the lid” release
occurs.
Cause of error /
measurements
• Weight difference in rotor components.
• Supporting lugs not lubricated.
• False IMBALANCE MODE is set (see chapter "Imbalance
Mode").
• Imbalance switch not connected.
• Imbalance switch is defective.
Measure on supply board / plug X3 (pin 4), see chapter
"Mains supply with supply board (A1)".
• Loose contact in cable or plug.
• Flat ribbon cable to control panel is defective.
Measure on control panel / plug X1 (pin 12)
• control panel or supply board is defective.
Error code reset Open the lid or perform a MAINS RESET.
CONTROL - ERROR 04
Error Lid lock is open during centrifugation.
Error
consequence
Slowing down until the ”open the lid” release occurs.
Cause of error
• Lid lock is defective and can be opened during
centrifugation.
• Loose contact in cable or in plug.
• Flat ribbon cable to control panel is defective.
• Control panel or supply board is defective.
Measurement Measure on supply board / plug X5 (pin 6) and on control
panel / plug X1 (pin 5).
Error code reset Perform a MAINS RESET.