EasyManuals Logo

Hettich EBA 21 Repair Instructions

Hettich EBA 21
42 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #24 background imageLoading...
Page #24 background image
24/42
6.5 Working hours
You can check and change the working hours only if the rotor is at standstill.
Keep the key pressed for 8 seconds.
After 8 seconds, SOUND / BELL XXX will be displayed.
Press the key again.
The working hours (CONTROL: ) of the centrifuge will be displayed.
To exit the working hours check screen, press the or key.
Press the
RCF
key to set the working hours.
Set the working hours with the and keys.
Press the
START
key to save the setting.
∗∗∗ ok ∗∗∗ is displayed for a short time to confirm that the setting has been saved.
6.6 Checking the motor slippage
The centrifuge control can readjust the speed by max. 5 % (control limit).
The error (N < MIN 13) is displayed if the rotor speed (ACTUAL speed) is lower than the
SET speed minus 5%.
Checking the slippage:
Start a centrifugation run and wait until the set speed is achieved.
Keep the key pressed (approx. 8 s) until the following is displayed:
X X X X
X X X X
________
________
Rotor speed Field speed
Slippage = (
field speed) - (rotor speed)
The slippage display automatically disappears after 8 seconds.

Table of Contents

Other manuals for Hettich EBA 21

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich EBA 21 and is the answer not in the manual?

Hettich EBA 21 Specifications

General IconGeneral
BrandHettich
ModelEBA 21
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals