5.3 Description and elimination of errors
TACHO - ERROR 01
Error During centrifugation the speedometer pulses are interrupted.
Error
consequence
The rotor slows down until it stops.
After the rotor stops, there is a DC braking for 30 sec.
An MAINS RESET during slowing-down causes a DC braking
for 3 min.
After the DC braking, the ”open the lid” release takes place.
Cause of error /
measurements
Speed sensor (speedometer) defective or loose contact on
plug. Measure speedometer pulses on plug X4 / supply
board (pin 4 - pin 2).
Flat ribbon cable to control panel, or flat ribbon cable to
frequency converter is defective.
Measure speedometer pulses on control panel / plug X1
(pin 14) and on frequency converter / plug S501 (pin 8).
Supply board or control panel or frequency converter is
defective.
Error code reset Open the lid. Turn the rotor by hand and perform a MAINS
RESET while the rotor is turning.
TACHO - ERROR 02
Error There are no speedometer pulses on the control panel after
start-up.
Error
consequence
The rotor slows down until it stops.
After the rotor stops, there is a DC braking for 30 sec.
An MAINS RESET during slowing-down causes a DC braking
for 3 min.
After the DC braking, the ”open the lid” release takes place.
Cause of error /
measurements
Start-up took place without the rotor.
Motor not connected.
Motor is defective.
Speed sensor (speedometer) defective, or loose contact on
plug. Measure speedometer pulses on plug X4 / supply
board (pin 4 - pin 2).
Flat ribbon cable to control panel, or flat ribbon cable to
frequency converter is defective.
Measure speedometer pulses on control panel / plug X1
(pin 14) and on frequency converter / plug S501 (pin 8).
No release signal to frequency converter.
Supply board or control panel or frequency converter is
defective.
Error code reset Open the lid. Turn the rotor by hand and perform a MAINS
RESET while the rotor is turning.

Table of Contents

Other manuals for Hettich EBA 21

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich EBA 21 and is the answer not in the manual?

Hettich EBA 21 Specifications

General IconGeneral
BrandHettich
ModelEBA 21
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals