3.8 Imbalance switch
A switch (break contact) detects imbalance.
Imbalance is detected only in running mode (starting, centrifuging and braking).
If imbalance is detected, the drive is brought to a standstill.
3.9 Safety devices
Mains input
Mains fuses
Radio interference
suppression filter
between supply board and frequency converter
Frequency converter
electronically protected
Motor
overtemperature switch > 135 °C
4 Troubleshooting procedures
Fuses in installation in which centrifuge is installed are intact.
Mains input fuses of centrifuge are intact.
Supply voltage present at (see circuit diagram):
Connecting cable
Appliance plug
Mains switch
Supply board, plug X5 (pin 1, pin 5)
Look for the displayed error code in the chapter "Error messages".
Remedy the error according to the instructions.
Carry out a functional check after every repair and whenever a component is
replaced, see chapter "Functional check after a repair".