EasyManua.ls Logo

Hettich EBA 21 - Page 15

Hettich EBA 21
42 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15/42
VERSION - ERROR 12
Error Differences in the initialization from control panel (EPROM) or
frequency converter.
Error
consequence
No further user operation is possible.
Cause of error
An incorrect EPROM has been plugged into control panel.
Centrifuge is not initialised. Carry out an initialization, see
chapter "Initialization".
Error code reset Perform a MAINS RESET.
N < MIN 13
Error Insufficient speed, motor slippage is too large.
The centrifuge control can readjust the speed by max. 5 %
(control limit).
This error is displayed if the rotor speed (ACTUAL speed) is
lower than the SET speed - 5%.
Error
consequence
The centrifuge slows down until the ”open the lid” release
occurs.
Cause of error
Motor is labouring (damage to bearings).
Motor has a short-circuited coil (coil is defective).
Loose contact in the electrical connections.
Frequency converter is defective.
Release signal to frequency converter was interrupted
during centrifugation.
Error code reset Open the lid and perform a MAINS RESET.
CONTROL - ERROR 21 - 26
Error Internal error in control panel.
Error
consequence
The centrifuge slows down until the ”open the lid” release
occurs.
Cause of error
Control panel is defective.
Error code reset Perform a MAINS RESET.