CP : control panel, FC : frequency converter, SB : supply board, CB : cooling board, CC : control cable, LL : lid locking, BC : braking chopper,
BR : brake resistor, MR : mains reset
15/49
ROTORCODE 10
Error cause An invalid rotor code was read during start-up.
Error
consequence
The centrifuge slows down until the ”open the lid” release
occurs.
Further cooling until NOMINAL temperature is reached.
Error remedy
• Magnetic coding on rotor is defective.
• Speedometer system is defective.
• Loose contact on speed sensor (speedometer) plug
• The rotation of the rotor (direction) is incorrect.
Measurement Also see section 10.
Error code reset Open the lid or perform an MR.
MAINS INTERRUPT
Error cause Interruption of mains supply during centrifugation.
Error
consequence
The centrifuge slows down until the ”open the lid” release
occurs.
− Switching on at the mains during centrifugation causes
slowing-down until the ”open the lid” release occurs.
− Switching on at the mains when the rotor has stopped
brings about the ”open the lid” release.
Error remedy
• Power supply has failed.
• Loose contact in electrical connections.
• CC to CP is defective.
Error code reset Open the lid and press the
START
key.
This error cannot
be reset by an MR
VERSION - ERROR 12
Error cause Differences in the initialisation from CP (EPROM) or FC.
Error
consequence
No further user operation is possible.
Error remedy
• An incorrect EPROM has been plugged into CP.
Measurement Also see initialisation section 5.2
Error code reset Perform an MR.