12.13. Appliance plug , overvoltage protection F1
Remove mains switch (see section 12.12 ).
Unscrew the two fastening screws at the appliance plug.
Pull the appliance plug out of the opening.
Remove the overvoltage protection from the appliance plug.
Replace appliance plug.
Unplug the cables at the overvoltage protection.
Replace overvoltage protection.
12.14. Imbalance switch S1
Remove plug from position X3 at the CB.
From below loosen the two fastening screws of the imbalance switch.
Loosen the fastening nuts from the motor. Lift up the motor and pick up the
imbalance switch through the opening at the centrifuge chamber.
Replace imbalance switch.
12.15. Lid lock
Remove the front panel.
Unplug all cable from the lid lock.
Loosen the to fastening screws on the to of the housing.
Exchange the lid lock.
12.16. Cooling board (CB) A3 (only UNIVERSAL 32 R)
Remove the front panel.
All plugs on the CB must be pulled out.
Unplug the two cables (RD and BK) at the SB.
Unscrew the four fastening screws at the CB.
Replace the CB.
12.17. Temperature sensor B1 in centrifuge chamber (only UNIVERSAL 32 R)
Remove plug from position X3 at the CB.
Remove the four bushes in the plug (unlock the bushes with appropriate tools at the
front of the plug). Then press out the four bushes.
Press out the temperature sensor in the centrifuge chamber.
Replace temperature sensor.
12.18. Temperature sensor B2 at condenser (only UNIVERSAL 32 R)
Remove plug number X2 at the CB.
Unscrew the temperature sensor (one screw).
Replace temperature sensor.