12. Assembling and disassembling components
Before assembling and disassembling components the following steps must be
executed:
Open the lid
Remove the rotor
Disconnect the centrifuge from mains
12.1. Front panel
Remove the three fastening screws at the lower edge of the front panel and
lift the front panel up.
Unplug all cable from the back side of the control panel.
12.2. Control board (CP) A4
Remove the front panel (see section 12.1 )
Remove the two fastening bows
Press the four fasting clips and press-out the control board.
exchange the control panel
12.3. EPROM at Control board
Remove the front panel ( see section 12.1 )
Pull the EPROM carefully out of IC-socket.
Pay attention to the polarity of the EPROM when installing.
Do not bend the IC-pins.
Before touching the EPROM ensure that your own static electricity is
discharged.
Replace the EPROM.
12.4. Speed sensor B3 (speedometer)
Remove the to fastening screws of the motor hood
Unscrew speed sensor (speedometer) from upper end plate of motor.
Remove the front panel ( see section 12.1 ).
Unplug plug number X4 from Supply board A1.
Remove the cable fixtures at the motor
Replace speed sensor (speedometer).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich UNIVERSAL 32 R and is the answer not in the manual?

Hettich UNIVERSAL 32 R Specifications

General IconGeneral
BrandHettich
ModelUNIVERSAL 32 R
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals