2.1 None of the Bed Functions Operate
Chapter 2: Troubleshooting Procedures
2
Hill-Rom® Basic Care™ Bed, Hill-Rom® 305 Manual Bed, Page 2 - 5
Hill-Rom® 405 Electric Bed Service Manual (MAN336 REV 2)
2.1 None of the Bed Functions Operate
1. All of the bed function lockouts are off.
If the indicator for a bed function lockout is on, the function is locked out.
Yes No
↓→Turn the bed function lockouts off. If this corrects the problem,
go to “Final Actions” on page 2-4.
2. Disconnect the bed from its power source and allow the bed to cool for 20
minutes. Connect the bed to its power source. The bed functions do not
operate.
Yes No
↓→Go to “Final Actions” on page 2-4.
3. The power cord is tightly seated in the connector on the control box.
Yes No
↓→Tightly seat the power cord into its connector on the control box.
If this corrects the problem, go to “Final Actions” on page 2-4.
Otherwise, go to step 4.
4. All of the control connectors are tightly seated in the control box.
Yes No
↓→Tightly seat the control connectors in the control box. If this
corrects the problem, go to “Final Actions” on page 2-4.
Otherwise, go to step 5.
5. Check the fuses in the control box at the head end of the bed. The fuses are
good.
Yes No
↓→Replace the fuses (refer to procedure 4.6 on page 4-14). If this
corrects the problem, go to “Final Actions” on page 2-4.
Otherwise, go to step 6.
6. Replace the control box (refer to procedure 4.1 on page 4-3). If this
corrects the problem, go to “Final Actions” on page 2-4. Otherwise, call
Hill-Rom Technical Support at 800-445-3720 (USA).