EasyManua.ls Logo

Hill-Rom P1441B - Page 31

Hill-Rom P1441B
176 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.3 The Head Down Function Does Not Operate
Chapter 2: Troubleshooting Procedures
2
Hill-Rom® Basic Care™ Bed, Hill-Rom® 305 Manual Bed, Page 2 - 9
Hill-Rom® 405 Electric Bed Service Manual (MAN336 REV 2)
8. Replace the controls that has the head down function that does not operate
(refer to procedure 4.21 on page 4-46). The head down function operates
from this siderail.
Yes No
↓→Replace the siderail with the faulty function (refer to procedure
4.21 on page 4-46). If this corrects the problem, go to “Final
Actions” on page 2-4.
9. Replace the control box (refer to procedure 4.1 on page 4-3). If this
corrects the problem, go to “Final Actions” on page 2-4. Otherwise, call
Hill-Rom Technical Support at 800-445-3720 (USA).

Table of Contents

Related product manuals