2.8 The Hilow Up Function Does Not Operate
Chapter 2: Troubleshooting Procedures
2
Hill-Rom® Basic Care™ Bed, Hill-Rom® 305 Manual Bed, Page 2 - 17
Hill-Rom® 405 Electric Bed Service Manual (MAN336 REV 2)
8. Replace the controls on the siderail that has the hilow up function that does
not operate (refer to procedure 4.21 on page 4-46). The hilow up function
operates from this siderail.
Yes No
↓→Replace the siderail that has the hilow up function that does not
operate (refer to procedure 4.21 on page 4-46). If this corrects
the problem, go to “Final Actions” on page 2-4.
9. Replace the control box (refer to procedure 4.1 on page 4-3). If this
corrects the problem, go to “Final Actions” on page 2-4. Otherwise, call
Hill-Rom Technical Support at 800-445-3720 (USA).