Features
TotalCare® Bariatric Bed and TotalCare® Bariatric Plus Therapy System 47
User Manual (143434 REV 8)
OPTI-REST
The OPTI-REST mode offers increased comfort to the patient while maintaining pressure relief. It vents the chest, seat,
and thigh zones producing a massaging wave-like action.
To Activate:
1. Press the Surface tab on the GCI.
2. Press the Start under OPTI-REST.
To Deactivate:
1. Press the Surface tab on the GCI.
2. Press Stop under OPTI-REST,
or
on the siderail, press Normal control.
Patient History
To View:
1. Press the Tools tab on the GCI.
2. Press View History.
3. Select the desired history to be viewed.
• Rotation: Displays the maximum number of
cycles/hour the patient has rotated and Hours,
Minutes in rotation, in 24 hours.
• Percussion: Displays the number of treatments
provided per 24-hour period.
• Vibration: Displays the number of treatments provided
per 24-hour period.
• OPTI-REST: Time spent in OPTI-REST mode since 12
AM.
• Head Angle: Time spent with the head of bed more
than 30° or 45° since 12:00 AM.
• Weight: Displays the weight increase or decrease in
24-hour periods.
• Chair: Time spent in Chair position since 12:00 AM.
• Bed Exit: Displays the time spent with the Bed Exit alarm on.
To Clear History:
1. Press the Tools tab on the GCI.
2. Press Clear (New Patient) next to Patient History.
It is recommended to clear all statistics between patients.
NOTE:
Clear History will clear all Therapy Statistics.