Features
54 TotalCare® Bariatric Bed and TotalCare® Bariatric Plus Therapy System
User Manual (143434 REV 8)
5. If the rotation therapy is not going to be used at the time of setting, press ENTER. Then release the
Graphical Caregiver Interface (GCI)® Control or select GO TO Home Screen for other options.
6. If not acceptable, scroll to Change Settings, Clear Changes, Cancel/Return, or go to Home Screen, press
ENTER.
To Stop Rotational Therapy:
Graphical Caregiver Interface (GCI)® Control method:
1. On the Home Screen, scroll to Normal/Standard. Press ENTER.
Siderail Method:
1. On the siderail opposite the Graphical Caregiver Interface (GCI)®
Control, press the Enable control and the Normal/Standard control.
Rotation Reminders:
• Rotation therapy will be suspended when:
• Any siderail is lowered. To restart rotation, raise siderail to the
upright position.
• Head of Bed (HOB) is raised higher than 40 degrees. To restart rotation, lower HOB.
• Foot of Bed (FOB) is lowered more than 30 degrees. To restart rotation, raise FOB.
• Chair position is attempted. To restart rotation, exit chair position.
• Percussion/Vibration, Max-inflate, or Turn Assist is active.
• The Therapy Suspended light on the siderail will blink when therapy has been suspended for any of the
above conditions.
• If CPR is activated, rotation therapy automatically stops and Max-inflate is activated. To resume rotation,
see “Rotational Therapy” on page 52.
• Use Alarm Silence (on the Graphical Caregiver Interface (GCI)® Control screen or the opposite siderail
panel) to turn off any audible alarms.
• Pulmonary Status is shown in the lower left corner. This shows: hrs/mins. rotated since 12 AM, active, off,
suspended, or removed.
• Check the Graphical Caregiver Interface (GCI)® Control screen if you are uncertain why the bed is
alarming. The reason will be displayed on the Graphical Caregiver Interface (GCI)® Control screen.
Percussion and/or Vibration Therapies
WARNING:
Relative Contraindication—Unstabilized spinal cord injuries. Under normal circumstances,
correct alignment on an air bladder system is easy to maintain. However, user error (such as
the activation of Turn Assist, Rotation, Opti-Rest, or Percussion/Vibration functions, when
available) or equipment malfunction (such as cushion deflation) could put correct alignment
at risk. The choice of a therapeutic support for such conditions as an unstabilized spinal cord
injury is based upon the medical judgment of professionals. Each case should be evaluated
individually.
Percussion is the clapping of the posterioral chest wall to loosen secretions in the lungs.
Vibration, or shaking, of the posterioral chest wall helps move lung secretions for easier removal.
The percussion and vibration therapies can be done separately or together as a sequential treatment.