2.12 Pulse Output Function
35
1
2
3
4
5
6
7
8
9
10
11
Measurement
Pulse Output
As soon as the function switch is set to pulse output,
the pulse voltage is output from the output terminal.
The initial settings are a frequency of 600 Hz and a
duty ratio of 50.00%. Turning off the instrument
restores these initial settings.
Setting Range Resolution Accuracy
Frequency 0.5, 1, 2, 5, 10, 15, 20, 25, 30,
40, 50, 60, 75, 80, 100, 120, 150,
200, 240, 300, 400, 480, 600, 800,
1200, 1600, 2400, 4800 Hz
0.01 Hz
± 0.005%rdg.
± 2 dgt.
DUTY ratio 0.39% to 99.60% 0.390625% ± 0.4%
Pulse width
1/ Frequency Range/ 256
± 0.01%rdg.
± 0.3 ms
Amplitude Fixed, 2.8 V ± 0.2 V
Output impedance: 3.5 k
Ω or less