4.11 Reference Impedance Setting
65
1
2
3
4
5
6
7
8
9
10
11
Power On Options
This procedure sets the reference impedance for the
decibel (dBm) conversion screen.
4.11 Reference Impedance Setting
34
1. In order to display the Power On
Option setting screen, hold dow
n
the SHIFT key while turning th
e
function switch.
2. Press the
REL key twice to displa
y
the Reference Impedance settin
g
screen.
3. Set the reference impedance.
RANGE key: Increases the value
DUAL key: Decreases the value
SHIFT key: Selects the digit
4. Confirm the selection by pressin
g
the Hz key.
5. Hold down the
SHIFT key to ent
er
the measurement screen.
Otherwise, turn the function switc
h
to OFF to finish setup.
1 to 9999 Ω
(default: 600 Ω)
1
SHIFT +
Power On
2
(twice)
5