56
4.2 Toggling the Percentage Display (4 - 20mA/0 - 20mA)
When measuring with the DCmA function, press the
DUAL key three times to display the percentage display in
the sub display.
This procedure sets the percentage display method.
4.2 Toggling the Percentage Display
(4 - 20mA/0 - 20mA)
4-20 mA Converts the current over a range of 4 mA to 20
mA into a percentage from 0 to 100%.
0-20 mA Converts the current over a range of 0 mA to 20
mA into a percentage from 0 to 100%.
34
1. In order to display the Power On
Option setting screen, hold dow
the SHIFT key while turning th
function switch.
2. Press the
HOLD key 5 times to di
play the Percentage Display se
ting screen.
3. Press the
RANGE key or the DUA
key to select the percentage di
play setting.
4. Confirm the selection by pressin
the Hz key.
5. Hold down the
SHIFT key to ent
the measurement screen.
Otherwise, turn the function switc
to OFF to finish setup.
4-20 mA (default)
/ 20 mA
1
SHIFT +
Power On
2
(5 times)
5
Select through
power on option
Percentage
conversion value =
Measured value[mA] - 4[mA]
16[mA]
X 100[%]
Measured value[mA]
20[mA]
X 100[%]
Percentage
conversion value =