Operating Precautions
9
For safety reasons, disconnect the power cord when the instrument is not used and before
connecting it to a device to be tested.
To avoid damage to the instrument, do not input the voltage to the output terminals. Also do
not short between any terminals.
When the instrument is turned off, do not apply voltage or current to the instrument. Doing
so may damage the instrument.
Do not connect or disconnect connection cords from the current sensor or 9555-10 Sensor
Unit while the instrument is turned on. Doing so may damage the instrument, current sen-
sor, or 9555-10 Sensor Unit.
Do not input current to the current sensor when it is not connected to the instrument or
when the instrument and 9555-10 Sensor Unit are turned off. Doing so may damage the
current sensor, instrument, or 9555-10 Sensor Unit.
Correct measurement may be impossible in the presence of strong magnetic fields, such as near trans-
formers and high-current conductors, or in the presence of strong electromagnetic fields such as near
radio transmitters.
To measure accurately
Warm up the instrument for more than 30 minutes before use.
To maintain sufficient measurement accuracy of the instrument, be sure to help heat radiation.
e.g.) Keeping away the instrument from a heat, leaving sufficient clearances around the instrument,
installing cooling fans to the rack in which the instrument is mounted, or other measures.
The recommended calibration interval of the instrument is one year.