3.6 Performing Synchronized Measurement with Multiple Instruments
(Multiple-instrument Synchronized Measurement)
81
You can perform synchronized measurement by connecting multiple (up to eight: one master and up to seven
slaves) instruments (PW3336/PW3337) with optional 9165 Connection Cord (BNC cable).
This functionality can be used to perform simultaneous measurement of multiple circuits by operating only the
instrument (PW3336/PW3337) set up as the master (instrument setting of OUT) and thereby controlling the
instrument (PW3336/PW3337) set up as the slave (instrument setting of IN).
The instrument (PW3336/PW3337) set up as the slave will match the timing of the instrument (PW3336/
PW3337) set up as the master for the following operations:
Internal calculations
Display updates
Data updates
Integration start, stop, and reset
Display hold
Zero-adjustment
•Key lock
3.6 Performing Synchronized Measurement with
Multiple Instruments (Multiple-instrument
Synchronized Measurement)
To avoid damaging the instruments, do not connect or disconnect cables while the instru-
ments are turned on.
Use a common ground for instruments (PW3336/PW3337) that are performing simultaneous
measurement. If different grounding is used, a potential difference will occur between the
master ground and the slave ground. Connecting the connection cable (for synchronization)
while there is such a difference may result in malfunction or damage.
Dedicated signals are used in synchronized measurement. Do not input signals other than
those designed for synchronized measurement. Doing so may result in malfunction or dam-
age.