3.2 Configuring Settings
55
Example: 1P2W×3
3.2.7 Setting the Timeout
Setting
Description
0.1 sec.
Use this setting when the frequency of the input set as the synchronization source is 10 Hz or higher. (De-
fault setting)
TIME OUT 1s/10s lamp off
1 sec.
Use this setting when the frequency of the input set as the synchronization source is less than 10 Hz.
TIME OUT 1s/10s lamp on
10 sec.
Use this setting when the frequency of the input set as the synchronization source is less than 1 Hz.
TIME OUT 1s/10s lamp on
2 Press to display the setting
screen shown to the left.
:
The screen varies depending on whether the wiring mode is
1P2W×2 or a mode other than 1P2W.
:
The screen varies depending on whether the wiring mode is
1P2W×3, 1P3W&1P2W and 3P3W&1P2W, 3P3W2M, or 3V3A
and 3P3W3M and 3P4W.
3 Press or to select the channel
to set.
Areas b, c, and d on the display correspond to CH1, CH2,
and CH3, respectively. The selected channel parameter will
flash.
4 Press or to set the timeout
(0.1 sec., 1 sec., or 10 sec.).
Settings: 0.1 → 1 → 10 ⋅⋅⋅
5 Set the timeout for other channels as
necessary.
6 Press to exit the settings.
The instrument will revert to the normal measure-
ment state.