1.4 Measurement Workflow
24
Configure the following settings as necessary:
Setting the frequency measurement range
See: "3.2.6 Setting the Frequency Measurement Range" (p.53)
Addressing display value variation: Displaying average measured values
See: "3.2.8 Displaying Measured Values as an Average (AVG: Averaging)" (p.57)
Measuring voltages in excess of 1,000 V: Using VT (PT) to make measurements
See: "3.2.9 Setting the VT and CT Ratio" (p.59)
Measuring currents in excess of 65 A: Using CT to make measurements
See: "3.2.9 Setting the VT and CT Ratio" (p.59)
When you wish to perform integration
See: "3.3 Integration" (p.62)
When you wish to measure harmonics
See: "3.4 Viewing Harmonic Measured Values" (p.71)
When you wish to measure efficiency
See: "3.5 Measuring Efficiency" (p.78)
When you wish to use the display hold, maximum value hold, and minimum value
hold functions
See: "3.10.1 Fixing Display Values (Display Hold)" (p.108)
"3.10.2 Displaying Peak, Minimum, and Maximum Values (Maximum Value Hold)" (p.109)
When you wish to use D/A output
See: "Examples of analog output"(p.98)
When you wish to use the RS-232C interface
See: "Setting the RS-232C communications speed"(p.119)
When you wish to use the LAN interface
See: "Setting the LAN’s IP address"(p.123)
When you wish to use the GP-IB interface
See: "Setting the GP-IB address"(p.133)
When you wish to perform synchronized measurement with multiple instruments
See: "3.6 Performing Synchronized Measurement with Multiple Instruments (Multiple-instrument Syn-
chronized Measurement)" (p.81)