3.2 Configuring Settings
52
4 Press or to set the
synchronization source.
Settings: U1 I1 U2 I2 U3 I3
DC ⋅⋅⋅
(Setting the source to DC will cause the SYNC. DC lamp for
the set channel to light up.)
5 Set the synchronization sources for the
other channels as necessary.
When using a wiring mode other than 1P2W, the synchronization source will be standardized using the
CH1 settings.
Setting the synchronization source to DC when using AC input will cause display values to fluctuate, pre-
venting accurate measurement. Set the synchronization source to a value other than DC when using AC
input.
The voltage and current for each channel will use the same synchronization source as set.
Input for the parameters selected as synchronization sources must be at least 1% of the range.
When the SYNC. ERROR lamp is lit up, accurate measurement cannot be performed.
Even when the SYNC. ERROR lamp is not lit up, accurate measurement cannot be performed if the syn-
chronization source input signal is in any of the following states:
1. When a signal with a frequency higher than the frequency measurement range (zero-crossing filter) is
being input
2. When a measured voltage or current of an AC component of an input signal is less than 1% of each
relevant range.
3. When a measured voltage or current of an AC component of an input signal is greater than 130% of
each relevant range.
4. When a signal with a frequency component within the frequency measurement range (zero-crossing
filter) other than the measurement signal’s fundamental wave frequency is superposed
Example: When the frequency measurement rage (zero-crossing filter) is 500 Hz, if a signal for which
zero-crosses occur at a frequency of less than 500 Hz (a modulation signal, noise, etc.) is
input on top of a 50 Hz input signal
The synchronization source cannot be changed while integration is being performed or during display hold
or maximum value/minimum value hold operation.
When using a timeout setting other than 0.1 sec. (1 sec. or 10 sec.) for channels with the 1P2W×2 or
IP2W×3, 1P3W&1P2W, and 3P3W&1P2W wiring modes, set the synchronization source to the voltage or
current for a channel whose timeout has been set to 1 sec. or 10 sec.
Example: If the CH1 and CH3 timeout setting is 0.1 sec. and the CH2 timeout is 10 sec. while using the
1P2W×3 wiring mode, set the CH2 synchronization source to U2 or I2.
6 Press to exit the settings.
The instrument will revert to the normal measure-
ment state.