6
2. SPECIFICATIONS
2.1 Measuring Performance
1) Electrical Resistance Measurement
SM-8213(RANGE : ×1 , ×10 , ×10
2
, ×10
3
, ×10
4
, or ×10
5
)
Measurement Range (MΩ) [RANGE=×10
R
]
Target Voltage
0.5 × 10
R
to 20 × 10
R
100 V
0.25 × 10
R
to 10 × 10
R
50 V
0.125 × 10
R
to 5 × 10
R
25 V
0.075 × 10
R
to 3 × 10
R
15 V
0.05 × 10
R
to 2 × 10
R
10 V
0.025 × 10
R
to 1 × 10
R
5 V
Accuracy of target voltage: ±3% of set value
Measuring output current: 50 mA maximum (Note: 1)
Accuracy of measurement: ±10% (A range of 10 times of the minimum
value of each range) (Note: 2)
SM-8215(RANGE : ×1 , ×10 , ×10
2
, ×10
3
, ×10
4
, or ×10
5
)
Measurement Range (MΩ) [RANGE=×10
R
]
Target Voltage
5 × 10
R
to 200 × 10
R
1000 V
2.5 × 10
R
to 100 × 10
R
500 V
1.25 × 10
R
to 50 × 10
R
250 V
0.5 × 10
R
to 20 × 10
R
100 V
0.25 × 10
R
to 10 × 10
R
50 V
Accuracy of target voltage: ±3% of set value
Measuring output current: 2 mA maximum (Note: 1)
Accuracy of measurement: ±10% (A range of 10 times of the minimum
value of each range) (Note: 2)
SM-8220(RANGE : ×1 , ×10 , ×10
2
, ×10
3
, ×10
4
, ×10
5
, ×10
6
, ×10
7
, or ×10
8
)
Measurement Range (MΩ) [RANGE=×10
R
]
Target Voltage
5 × 10
R
to 200 × 10
R
1000 V
2.5 × 10
R
to 100 × 10
R
500 V
1.25 × 10
R
to 50 × 10
R
250 V
0.5 × 10
R
to 20 × 10
R
100 V
0.25 × 10
R
to 10 × 10
R
50 V
0.125 × 10
R
to 5 × 10
R
25 V
0.05 × 10
R
to 2 × 10
R
10 V
Accuracy of target voltage: ±3% of set value
Measuring output current: 2 mA maximum (Note: 1)
Accuracy of measurement: ±10% (A range of 10 times of the minimum
value of each range) except for×10
8
range. (Note: 2)
±20% for×10
8
range