EasyManua.ls Logo

Hioki SM-8213 - Self-Calibration 1; Self-Calibration 2

Hioki SM-8213
58 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
21
4) Confirm that the accessory shorting plug is inserted into HV-EN
connector on the rear the unit.
5) Leave the Rx /+ measuring terminals free at this stage – do not connect
the measuring rods and the sample to be measured to the terminals.
6) Set the RANGE selector switch to the ×1 position – the minimum
Multiplier.
7) Set the VOLTAGE SELECTOR switches to the lowest voltage available.
8) Turn the POWER switch ON.
Allow the unit to warm for a minimum of 30 seconds to obtain the
specified performance. However, a measurement can be done after a
warm-up of several seconds.
5.2 Self-calibration 1
1) Set the RANGE switch to the CAL position. (The display will show CAL1.)
The calibration will be verified by the internal calibration circuit. Verify
that the measured data display indicates a value within the range of .970
to 1.03.
2) If the value falls within the normal range, the self-calibration is complete.
Set the RANGE switch to a position other than
CAL. (If self-calibration 1
yields a result within the normal range, there is no need to perform
self-calibration 2.)
5.3 Self-calibration 2
1) Press the SELECT switch while CAL1 is being displayed. (The display will
show CAL2; see “4-4 Measurement Condition Setting Display” for more
information.)
2) Increase the RANGE switch to the
×10
2
range.
3) Verify that the screen is displaying values with 5 digits and press the
SELECT switch.
4) This completes self-calibration 2. Repeat self-calibration 1 and verify
that the value falls within the proper range.

Table of Contents

Related product manuals