Chapter 5: Device operation 59
Parameter screen
The ventilation and alarm parameters are
shown on the parameter screen. The phy-
sician can create up to 3 sets with differ-
ent ventilation modes and parameters.
PCV
Set 1
75%
50
40
30
20
10
0
p 19.8 hPa f
12 bpm V
I
0.570 lI:E 1:1.5
Set 3Set 2
Time Inspiration
I-Slope
E-Slope
Sigh Function
High Inspiration Volume
Low Inspiration Volume
High Minute Ventilation
2.0 s
3
3
Off
1.00 l
0.20 l
Off
Set 1
A
B
C
A Active set B Ventilation parameters
C Alarm parameters
Figure 21: Parameter screen
Activating another ventilation set
Risk of injury due to activating the wrong ventilation set!
The ventilation sets can contain different ventilation and alarm
parameters. Sets are not suitable for all applications.
Ö
Only activate the sets that the doctor has discussed with you.
1. On the home screen, navigate to “Parameters” and press the MFK.
2.
Navigate to the ventilation set you
wish to activate.
PCV
75%
50
40
30
20
10
0
p
hPa f
bpm V
I
lI:E
Set 3
Activate Set
Mode
IPAP
PEEP
Frequency
Time Inspiration
I-Slope
No
PCV
20.0 hPa
5.0 hPa
12 bpm
2.0 s
3
Set 1 Set 2
3. Press the MFK twice. This brings you
to the set to activate it.
PCV
Set 1
75%
50
40
30
20
10
0
p
hPa f
bpm V
I
lI:E
Set 3
Activate Set
Mode
IPAP
PEEP
Frequency
Time Inspiration
I-Slope
Set 1 Set 2
No
PCV
20.0 hPa
5.0 hPa
12 bpm
2.0 s
3