88 Chapter 9: Appendix
Technical data
Mains operation 100 to 240 V AC (-20%, +10%), 50 to 60 Hz
DC operation
24 V DC / 3.75 A
Internal battery operation
Lithium-ion battery, 15 V (nominal voltage) / 3.2 Ah
Max. power consumption 75 W
Electrical protection class Protection class II
Specifications and performance characteristics
Dimensions (W x D x H) 215 x 203 x 115 mm
Weight 2.2 kg
Max. inspiratory working pressure
1
LAVI 30: 30 hPa
LAVI 40: 40 hPa
Min. inspiratory working pressure
1
4 hPa
Supply voltage
Operating conditions
Temperature range + 5°C to + 40°C
Relative humidity 10% to 95%, non-condensing
Air pressure range
3
600 hPa to 1100 hPa (approx. 4000 to -400 m)
1
Ensured by pressure control.
2
AquaTREND uni
3
The air flow rate decreases with increasing altitude.
Transport conditions
Temperature range
With battery
Without battery
- 20°C to + 50°C
- 25°C to + 70°C
Relative humidity 5% to 90%, non-condensing
Air pressure range
1
250 hPa to 1100 hPa (approx. 10000 to -400 m)
Storage conditions
Store in a dry, vibration-free place, in an upright posi-
tion; store device and accessories in their original
packaging