66 Chapter 6: Alarms and messages
Alarm test
The user does not have to perform any alarm tests. However, the following describes
an option to call the “Disconnection” alarm in order to perform a manual check of
the alarm system.
To check the “Disconnection” alarm:
1. Connect the device to the mains supply.
2. Switch the device on. Do not connect the leakage tube system!
3. Start ventilation.
4.
After a few seconds, the “Disconnection”
alarm occurs. If this is not the case, the device
must be returned for servicing.
PCV
Set 1
75%
50
40
30
20
10
0
p
19.8
hPa f
12
bpm V
I
0.570
lI:E
1:1.5
Statistics
Parameter Event Log
Service
Monitoring
Thursday I 19.05.2016 I 23:10
System
Disconnection