Chapter 9: Appendix 95
Guidance and manufacturer’s declaration - electromagnetic immunity
Immunity tests IEC 60601 test level Compliance level Electromagnetic envi-
ronment - guidance
Voltage dips, tem-
porary interruptions
and voltage fluctua-
tions in the voltage
supply lines as per
IEC 61000-4-11
> 95 % dip of U
T
for
1/2 period
> 95 % dip of U
T
for 1
period
> 30 % dip of U
T
for
25 (50 Hz) periods/
30 (60 Hz) periods
> 95 % dip of U
T
for
250 (50 Hz) periods /
300 (60 Hz) periods
> 95 % dip of U
T
for
1/2 period
> 60 % dip of U
T
for
5 periods
> 30 % dip of U
T
for
25 periods
> 95 % dip of U
T
for
5s or 6s
The quality of the sup-
ply voltage should cor
-
respond to one of the
typical business or hos-
pital environments. If the
user of the LAVI venti-
lator requires continued
functioning even in the
event of an interruption
to the power supply, we
recommended supplying
the LAVI ventilator using
an uninterruptable power
supply (UPS) or a battery.