EasyManua.ls Logo

Hologic DISCOVERY C - Page 81

Hologic DISCOVERY C
274 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Discovery QDR Series Technical Manual
3-29
8. Click X-Rays (F2) to turn X-Rays off.
9. Change the X-ray Mode to 3.
Figure 3-17. Peak Potential Mode 4
10. Click X-Rays (F2) to turn X-Rays on.
11. Observe the oscilloscope. You should see a trace similar to Figure 3-18, alternat-
ing pulses, approximately 4ms in duration, with a peak amplitude of 14V and
10V respectively (corresponding to 140kVp and 100kVp inside the tank).
Figure 3-18. Peak Potential Mode 3
If the scope trace seen is not as shown in either Figure 3-17 or Figure 3-18, the system
may have insufficient AC power, a faulty X-ray Controller, or defective Tank.
Note: Enter this information into Enter this information into the Radiation Measurement
Report (see example on page 3-70).
AW-00336_001-0318
AW-00336_001-0319

Table of Contents

Related product manuals