Model
8754A
Calibration
Calibrate
the
~754A/8748A
combination
as follows:
S21
-
Forward
Transmission
S-Parameter Measurements
Connect
"through."
Set Test Set:
S11
/S2
1.
Perform Insertion Loss
or
Gain Calibration, B/R, using Channel 2, see Page 18.
Perform Insertion Phase Calibration, PHASE B/R, using Channel 2, see Page 19,
but
do
not
adjust Channel 2
OFFSET
vernier.
S12
- Reverse
Transmission
Set Test Set:
S22
/S12
.
Verify
that
both
B/R
and
PHASE
B/R
measurements
remain
near
0 dB, for
both
for-
ward
and
reverse.
Any
offsets
between
forward
and
reverse
must
be
noted
and
manually
subtracted
from
the
measured value. A phase-length difference
between
forward
and
reverse
will
occur
if
the
connections
from
ports
1
and
2
to
the
test
device are
not
the
same length.
S
11
-
Input
Reflection
Connect
short
circuit
to
Port
1
measurement
plane.
Set Test Set:
S11
/S21'
Perform
Reflection
Coefficient
Calibration,
POLAR
AIR,
see
Page
23,
but
do
not
ad-
just
CHANNEL
2
OFFSET
vernier.
S22
-
Output
Reflection
Connect
short
circuit
to
Port
2
measurement
plane.
Set Test Set:
S22
/S12'
Verify
that
trace
remains
near
outer
circle, ± 180° for
both
forward
and
reverse.
Any
off-
sets
between
forward
and
reverse
must
be
noted
and
manually
subtracted
from
the
mea-
sured
value. A phase-length difference
between
forward
and
reverse will
occur
if
the
connec-
tions from
ports
1
and
2
to
the
test
device are
not
the
same length.
27