Installation and Setup
Specifications
A-11
1Specifications
Defibrillator
Waveform: Damped sinusoidal (Lown).
Output Energy (Delivered): 2, 3, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 300, and
360 joules.
Charge Control: Push-button on apex paddle and on front panel.
Charge Time: Less than 5 seconds to 360 joules with battery present. Less than 15
seconds to 360 joules on AC only.
Armed Indicators: Charge done tone, charge done lamp on apex paddle, and
available energy indicated on display.
Paddle Contact Indicator: 3-color LED bar graph array on STERNUM paddle
indicates quality of defibrillator paddle contact before discharge.
Paddles: Standard paddles are anterior/anterior, adult and pediatric. Adult
electrodes (83 cm sq) slide off to expose pediatric electrodes (21 cm sq). Paddle cord
is 10 ft (3 m). Full range of internal paddles available.
Synchronizer: SYNC message appears on monitor and is annotated periodically on
recorder while in synchronous mode. An audible beep sounds with each detected R-
wave, while a marker on the monitor and sync designator on the recorder strip
indicate the discharge point.
Environmental Operating Conditions: 0 to 55 deg C, 15 to 95% relative humidity,
15,000 ft altitude.
ECG Trace SWEEP, SCROLL ECG trace style.
Table 1-2 Setup Menu 2 Settings
Setting Choices Description