GV10x OPERATION AND SAFETY MANUAL
B.3. Inspection of System Equipment (GV10x Source and Controller)
If Source is left mounted to a vacuum chamber, observe that system base
pressure is reached. This will insure that the GV10x is not causing leaks.
B.4. Safety Interlocks
The GV10x Source employs an interlock that terminates/prevents RF power whenever the
Source control cable is NOT connected. The same applies to the RF coax cable.
The interlock as shown in Figure B-1 is located inside the
GV10x Source enclosure. The general location is marked
in the photo with a yellow arrow.
NOTE: Only a technician authorized by ibss Group
should ever remove the Source cover.
RF exposure at cable and equipment contact points.
Will activate if either control cable or RF coax cable are
disconnected at either end.
Equipment Condition
After Interlock Activated:
Although electrical power is still going to the Controller,
the Source and Controller are in the “Plasma Off (STOP)”
state. This is a HARD STOP in that the user cannot
restart by pressing ON at the Controller or in the Remote
Control Software panels (the RF Generator is powered
OFF).
Figure B-1. Source Interlock