Rev 2015-03-02 31
GV10x OPERATION AND SAFETY MANUAL
Appendix C: GV10x Remote Control Specifications
COM port settings are 9600,8,N,1,N. Transmission characters to the GV10x are case sensitive
and spaces are not permitted. Commands begin with non-numeric characters. All messages IN
& OUT end with a ‘carriage return’ ‘linefeed’ pair (0x0D 0x0A). This specification applies to
PLC program version GVP101217.1210.2.ezc and later. NOTE: See Table C.2 for new
commands supported by PLC program version GVP141230.0000.4.ezc and later.
TABLE C-1 API for GV10x Controllers Manufactured Prior to Jan 15, 2015
P1 (P for Power, 1 for ON)
P0 (P for Power, 0 for OFF)
R1 (R=RF at Power-Up , 1=ON)
R0 (R=RF at Power-Up, 0=OFF)
R0 (RF OFF at Power-Up [default])
Wmn (W for Watts, m=0-9, n=0-9)
Wmn (W for Set Watts, m=0-9, n=0-9)
Hmn (H for Hours, m=0-9, n=0-9)
Hmn (H for Set Hours, m=0-9, n=0-9)
Mmn (M for Minutes, m=0-5, n=0-9)
Mmn (M for Set Minutes, m=0-5, n=0-9)
Smn (S for Seconds, m=0-5, n=0-9)
Smn (S for Set Seconds, m=0-5, n=0-9)
Wmn (W for Watts Set, m=0-9, n=0-9)
Hmn (H for Hours Set, m=0-9, n=0-9)
Mmn (M for Minutes Set, m=0-5, n=0-9)
Smn (S for Seconds Set, m=0-5, n=0-9)
Vm.n (V for Volts, m=0-9,n=0-9)
T (read Chamber Pressure in Torr)
Tnnn (T for Torr, nnn= 1 to 999, if >= 1T)
Tm-e (T for Torr, m=1-9, e=1-5, if < 1T)
Examples:
T 4 (pressure is 4T)
T7-3 (pressure is 0.007T)
F (read PLC & Panel program versions)
GVP060209@1115a.ezc
GVT060209@1115a.ezx
I (read hardware IDs: Controller &
Source Serial Numbers)
GetL (Get Lost Time, no plasma)