EasyManua.ls Logo

iDig 2D Touch - Page 181

iDig 2D Touch
187 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration
181
Machine Element Lengths are Way Off After Calibration (cont.)
Extension poles not used:
Make sure that the elevation difference between the low & high laser pointer levels have been measured accurately AND entered into
the Control Box correctly. Make sure the length units are correct as well.
Sensors:
Check in the Sensor Check menu if a sensor angle value freezes from time to time.
Magnetic target tube:
Make sure the magnetic target tube got attached to the correct bucket pivot point.
Definition:
A variance of 10% or less of the calculated element lengths compared to the actual measurements is “a little bit off.
An inaccuracy of ± 0.4 inches (10 mm) of each element length is acceptable and won’t effect the system accuracy.
If the machine has a 2
nd
boom:
Its important to change the angle between the main boom & 2
nd
boom significantly after each laser pointer shot without repetition of
certain angular orientations.
Laser pointer:
The units laser dot should NOT blink during the calibration procedure.
Check the units calibration to make sure that the laser dot is really horizontal.
Level the point laser very precisely using the bubble vial on the big silver adaptor or on the tripod.
If there’s heavy wind that causes the laser dot to bounce up & down, especially at the 2
nd
level, try a setup without the tubes using the
metal frame inside the cab next to the door as your higher level setup. Use the magnetic L-Bar to attach the laser pointer.
Machine Element Lengths are a Little Bit Off After Calibration