EasyManua.ls Logo

iDig 2D Touch - Page 28

iDig 2D Touch
187 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation Procedures Bucket Calibration
Automatic Bucket Calibration - Dog Bone
You will require a stable point on which to place the tip/edge of the bucket.
This point needs to be high enough so that you can open and close the bucket as much as possible.
The point needs to not move, deflect of change position when the bucket is placed on it.
Be careful with curved dog bones since they are extremely difficult to calibrate depending upon where the sensor is placed.
Ideally you should weld a plate onto the dog bone so that the center of the Sensor is in line with the two pivot pins that
connect to the dog bone.
The pop-up screen will remind you to take care when
placing the tip/edge of the bucket to keep on the
EXACT same point during the calibration process.
Press “OK” and then “Start.
Select “On the Dog Bone” to commence the
automatic calibration process.
28